AM16/32B Relay Analog Multiplexer
Multiplexer to Datalogger wiring - Signal lines from two COM
terminals are connected to two consecutive single-ended analog input
channels. One COM terminal is connected to a datalogger switched
excitation channel, and the remaining COM line connects to
datalogger ground. Up to 32 potentiometers may be measured by two
single-ended datalogger channels.
6.3.3 Four Wire Half Bridge (Measured Excitation Current)
Sensor to Multiplexer Wiring - one sensor per input SET. The panel
switch is set to “4x16” mode.
Multiplexer to Datalogger Wiring - One COM line is tied to a
datalogger excitation channel, and two COM lines to a differential
analog input. The remaining COM line is connected to the H side of a
datalogger differential channel along with a fixed resistor. The other
side of the resistor connects to the L side of the differential channel
and to ground (Figure 15). Up to 16 four wire half-bridges may be
measured by two differential datalogger channels in this manner.
/
0
"4 X 16" Mode
COM H (ODD)
COM L
COM H (EVEN)
COM L
ODD H
ODD L
EVEN H
EVEN L
SHIELD SENSOR SHIELD
S
COM
CR23X/
CR800/
CR850/
CR1000
21X/
CR7
CR10(X)
CR3000/
CR5000
EX E E VX
H H H H
L L L L
AG
H H H H
L L L L
G
FIGURE 15. Four Wire Half Bridge Hook-up and Measurement
The CR5000 and CR3000 also have current excitation channels which
allow a resistance measurement. Because the excitation current is
known, it is not necessary to measure the voltage across a fixed
resistor to determine the current as in Figure 15. See Section 5.3 for
an example.
29