MetPak User Manual Doc No. 1575-PS-0001 Issue 2 Gill Instruments Limited Saltmarsh Park, 67 Gosport Street, Lymingto

Strona 2

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 4.3 Aspiration Shield Secification for model 41003 i

Strona 3 - Contents

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 5 INSTALLATION REQUIREMENTS Host system  Access to th

Strona 4 - 2 INTRODUCTION

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 6 INSTALLATION Do NOT remove the black “rubber” transd

Strona 5 - 3 PRINCIPLE OF OPERATION

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 6.3 Electrical Cable Cable type Generic description –

Strona 6 - 3.2 Radiation Shield

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 Connecting the Analogue Output (Option 3 and 4) Remove

Strona 7

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 TB3 TB1 Connections from Win

Strona 8

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 Connections to WindSonic cable for MetPak options 5 and

Strona 9 - 4.2 Specification

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 7 MECHANICAL Orientation Normally the MetPak is mounte

Strona 10 - 4.5 Complete Assmebly

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 Figure 4 Alignment & Mounting details 17

Strona 11 - 5 INSTALLATION REQUIREMENTS

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 8 OUTPUT FORMATS On applying power to the MetPak, it wi

Strona 12 - 6 INSTALLATION

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 1

Strona 13 - 6.3 Electrical

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 WindSonic node address The default setting is ‘Q’ do no

Strona 14 - Analogue output wind

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 9 ANALOGUE OUTPUTS AND OPTIONS Note : Analogue outp

Strona 15

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 9.1 Entering Configuration mode (see section 8.1 for m

Strona 16 - MetPak Option 6 Analogue

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 9.6 Analogue settings The following commands are only v

Strona 17 - 7 MECHANICAL

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 9.7 Configuration / Diagnostic Information Each of th

Strona 18

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 10.1 SDI-12 COMMANDS SDI-12 configured units comply wit

Strona 19 - 8 OUTPUT FORMATS

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 11 MAINTENANCE & FAULT-FINDING 11.1 Cleaning If the

Strona 20 - (not displayed)

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 11.4 Returning unit If the unit has to be returned, it

Strona 21

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 12 TESTS 12.1 Bench test 1. Couple the MatPak to the ho

Strona 22 - 3 ENTER

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 13 APPENDICES 13.1 Glossary & Abbreviations Item

Strona 23 - 9.6 Analogue settings

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 Contents Contents...

Strona 24 - 10 METPAK SDI-12

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 29 13.3 Electrical Conformity EC DECLARATION OF CONFORMI

Strona 25 - 10.1 SDI-12 COMMANDS

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 13.4 Model 41003 Multiplate Radiation Shield Manual 3

Strona 26 - 11.3 Fault-finding

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 31

Strona 27 - 11.5 Status (error) codes

13.5 Rotronic HyrgoClip S3 manual 13.6 Campbell CS215 Temperature and Humidity PROBE manual

Strona 28 - 12 TESTS

CS215 Temperature and Relative Humidity Probe User Guide Issued 8.6.06 Copyright © 2004 Campbell Sc

Strona 29 - 13 APPENDICES

GuaranteeThis equipment is guaranteed against defects in materials andworkmanship. This guarantee applies for twelve months from date ofdelivery. We w

Strona 30 - 13.3 Electrical Conformity

Contents 1. Introduction ...1 2. Specifications...

Strona 31

Figures 1. Screen Capture of SDI-12 Transparent Mode on Edlog Array-Based Datalogger (CR10X) using Control Port 8 and Changing SDI-12 Address from 0

Strona 32

CS215 Temperature and Relative Humidity Probe The CS215 Temperature and Relative Humidity probe is designed as a reasonably priced probe for general m

Strona 33

CS215 Temperature and Relative Humidity Probe Response Time with Filter: <15s (63% response time in air moving at 1 m/s at humidities <85% ) E

Strona 34 - Relative Humidity Probe

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 1 FOREWORD Thank you for purchasing the MetPak manufact

Strona 35 - Guarantee

User Guide For other applications the CS215 can be installed in any orientation. Install it away from obvious sources of heat, including solar radiati

Strona 36

CS215 Temperature and Relative Humidity Probe • If long cabling or interference appear to be causing spurious readings then use the `checksummed’ ve

Strona 37 - Figures

User Guide NOTE: The sensor will automatically switch to making measurements every 11 seconds (+/-2 second) the first time it receives the aR! comma

Strona 38 - Humidity Probe

CS215 Temperature and Relative Humidity Probe ;{CR10X} ;Example program for two CS215 sensors on control port 1 ;Measures every minute and stores som

Strona 39 - 3. Installation

User Guide 5.1.2 Example CRBasic Program Measuring One Sensor Every 30 Seconds The following example is written for the CR200 datalogger. Programs f

Strona 40 - 4. Wiring

CS215 Temperature and Relative Humidity Probe The aR! Command would therefore be preferable in fast running programs as it takes less time to execute

Strona 41 - 5. Reading the Sensor

User Guide CallTable CS215dat 'Call the table instruction to calculate and store the Avs NextScan EndProg 6. Changing the SDI-12 Address Usi

Strona 42

CS215 Temperature and Relative Humidity Probe 7. To change the SDI-12 address, enter the command aAb!; where a is the current address from the above

Strona 43

User Guide 5. To activate the SDI-12 transparent mode on Control Port p, enter ∗8. The TD datalogger will respond with a “∗” prompt. At the “∗” prom

Strona 44 - 5.2.1 Edlog Dataloggers

CS215 Temperature and Relative Humidity Probe 6.3 CR200 Series Datalogger 1. Connect a single CS215 to the datalogger using Control Port C1/SDI12 a

Strona 45 - 5.2.2 CRBasic Dataloggers

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 3 PRINCIPLE OF OPERATION 3.1 Wind Speed and Direction

Strona 46 - EndProg

User Guide 6.4 CR1000 Datalogger The CR1000 datalogger supports changing the address in a similar way to the CR200. In terminal mode issue the SDI12

Strona 47

CS215 Temperature and Relative Humidity Probe • The correct orientation is with the black moulded tip of the element (that contains the sensing comp

Strona 48

User Guide Sensor connector sticking out of the end of the tubeGold pins Sensing Element Sensing part of the element. Printing on this side. Centre li

Strona 49 - 6.3 CR200 Series Datalogger

A-1 Appendix A. Environmental Performance This Appendix details tests and limitations of the sensor when exposed to extremes of the environment. A.1

Strona 50 - 6.4 CR1000 Datalogger

CS215 Temperature and Relative Humidity Probe A-2 When used outside the range of normal conditions or when subject to prolonged periods of condensati

Strona 51

B-1 Appendix B. Sensor Element Calibration The sensor elements are individually calibrated during manufacture. The manufacturer publishes the certif

Strona 52

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com info@campbellsc

Strona 53 - Performance

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 Figure 2 Compass Points 3.2 Radiation Shield The M

Strona 54 - A.4 Measurement below 0˚C

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 3.3 Temperature and Relative Humidity Sensor Two t

Strona 55

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 4 OPTIONS AND SPECIFICATION 4.1 Options and Par

Strona 56

MetPak Doc No 1575-PS-0001 Issue 02 CN 4240 March 2006 Option 6 1575-PK-006 WindSonic for MetPak