
CS450/CS455 Pressure Transducer
5.2 RS-232
TABLE 5-2. RS-232 Wiring
Color
CS450/CS455
Function
Connection
RS-232 9-pin
Red +12vdc Power Source
Black Power Ground Power Ground
White RS-232 Tx (Output) Transmit Pin 2 Rx (Input)
Blue RS-232 Rx (Input) Receive Pin 3 Tx (Output)
Yellow Digital Ground Ground Pin 5 GND
Clear Shield GND Ground
5.3 A150/CWS900
The A150 desiccated case allows the CS450/CS455 to be connected to the
CWS900 wireless sensor and then used in a wireless network. FIGURE 5-1
shows the CS450 wired to the A150’s terminal block. The 26972 cable is
included with the A150.
BLUE
RED
CLEAR
GREEN
1
2
BLACK
3
4
WHITE
5
6
BLACK
#26972 Cable
#19520
CWS900 Connector
WHITE
BLUE
BLACK
YELLOW
CLEAR
RED
TUBE
N.C.
N.C.
Terminates in A150 Enclosure
150P
essu
e Senso
CS450
Sensor TX (Out)
Sensor RX (In)
Power GND
Digital GND
Shield
Power
Vent
(Pin #)
FIGURE 5-1. A150 wired for CS450 pressure sensor and CWS900
connector
6. Programming
6.1 Using SCWIN
SCWIN is the easiest and typically the preferred method for programming the
datalogger. SCWIN generates a wiring diagram that shows how to connect the
CS450/CS455 to your Campbell Scientific datalogger.
The sections that immediately follow are for CRBasic and Edlog.
NOTE
16
Komentarze do niniejszej Instrukcji