Strona 1 - HC2S3 Temperature

HC2S3 Temperatureand Relative Humidity ProbeRevision: 12/11Copyright © 1990-2011Campbell Scientific, Inc.

Strona 2

Model HC2S3 Temperature and Relative Humidity Probe The black outer jacket of the cable is Santoprene® rubber. This compound was chosen for its resis

Strona 3 - Warranty

Model HC2S3 Temperature and Relative Humidity Probe shield (please refer to the 43502 product manual for details). The 41003-5 Radiation Shield has a

Strona 4 - Assistance

Model HC2S3 Temperature and Relative Humidity Probe CM200 Series CrossarmFIGURE 3-2. HC2S3 and 41003-5 Radiation Shield on a CM200 Series Crossarm

Strona 5 - HC2S3 Table of Contents

Model HC2S3 Temperature and Relative Humidity Probe TABLE 4-1. Connections for Single-Ended Measurements Color Description CR1000, CR3000, C

Strona 6 - Figures

Model HC2S3 Temperature and Relative Humidity Probe 8 5. Example Programs This section is for users who write their own datalogger programs. A datal

Strona 7 - Humidity Probe

Model HC2S3 Temperature and Relative Humidity Probe TABLE 5-3. Wiring for Single-ended Measurement Examples Color Description CR1000 CR10(X) Brown

Strona 8 - 2. Specifications

Model HC2S3 Temperature and Relative Humidity Probe CR10(X) program using single-ended measurement instructions ;{CR10X} program to measure HC2S3 wit

Strona 9

Model HC2S3 Temperature and Relative Humidity Probe 10: End (P95) 11: If time is (P92) 1: 0 Minutes (Seconds --) into a 2: 60 Interval (

Strona 10 - 3. Installation

Model HC2S3 Temperature and Relative Humidity Probe BeginProg Scan (1,Sec,1,0) 'Run main scan 1 second Battery (Battery_volts) PanelTem

Strona 11

Model HC2S3 Temperature and Relative Humidity Probe CR1000 program using differential measurements 'CR1000 program to measure HC2S3 with differen

Strona 13

Model HC2S3 Temperature and Relative Humidity Probe 4: If (X<=>F) (P89) 1: 4 X Loc [ RH ] 2: 4 < 3: 103 F 4: 30

Strona 14 - 5. Example Programs

Model HC2S3 Temperature and Relative Humidity Probe voltage drop (Vd), along the signal reference/power ground, is given by Eq. (1). )' (1000 m

Strona 15

Model HC2S3 Temperature and Relative Humidity Probe 16 3. Verify the green power wire is connected to the 12V, SW12V, or 5V terminal. When SW12V is

Strona 16

Appendix A. Absolute Humidity The HC2S3 measures the relative humidity. Relative humidity is defined by the equation below: 100 eeRHs∗= (A-1) whe

Strona 17

Appendix A. Absolute Humidity A-2 CR1000 Program that Computes Vapor Pressure and Saturation Vapor Pressure 'CR1000 program that calculates Vapo

Strona 18 - Instructions

Appendix B. Changing the HC2S3 Settings B.1 HC2S3 Default Settings The HC2S3 probe has the following default settings, which can be changed as descr

Strona 19

Appendix B. Changing the HC2S3 Settings B.2 Software and Hardware Requirements For temperature (Analog Output 2), the HC2S3 default range is -40 to

Strona 20

Appendix B. Changing the HC2S3 Settings Click on “Device Manager”, select “Analog Outputs” to see the following screen: Change the lower and upper r

Strona 21 - 8. Troubleshooting

Appendix B. Changing the HC2S3 Settings B-4 Example measurement instructions for CR1000 datalogger, with the sensor wired to SE channel 2: Public Air

Strona 22 - 9. References

Appendix C. HC2S3 Digital Communications C.1 HC2S3 Digital Interface Specifications The HC2S3 has a UART (Universal Asynchronous Receiver Transmitte

Strona 23 - ∗= (A-1)

Warranty “PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in materi

Strona 24 - A.1 Measurement Below 0°C

Appendix C. HC2S3 Digital Communications Command Format: { ID Adr RDD Chksum or } CR Answer format: { ID Adr RDD Chksum or } C

Strona 25 - Settings

Appendix C. HC2S3 Digital Communications C.3 RS-485 Communications using an MD485 RS-485 Interface The HC2S3 can be interfaced to a CSI datalogger t

Strona 26

Appendix C. HC2S3 Digital Communications Sensor Wiring: E2-05XX-MOD Cable MD485 CR1000 Blue A Red B Green 12V Gray/Yellow G Cl

Strona 27

Appendix C. HC2S3 Digital Communications C.4 RS-485 Communications using an SDM-SIO1 Serial I/O Module The HC2S3 can be interfaced to a CSI datalogg

Strona 28

Appendix C. HC2S3 Digital Communications C-6 BeginProg SerialOpen (SensorPort,19200,51,100,200) '51 is for half duplex String_1 = "|{

Strona 30

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbel

Strona 31 - Interface

Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing

Strona 32

HC2S3 Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to

Strona 33 - Serial I/O Module

HC2S3 Table of Contents ii C. HC2S3 Digital Communications...C-1 C.1 HC2S3 Digital Interface Specifications ...

Strona 34

Model HC2S3 Temperature and Relative Humidity Probe 1. General Description The HC2S3 is a digital probe with 0 to 1 V linear output signals for tempe

Strona 35

Model HC2S3 Temperature and Relative Humidity Probe The HC2S3 ships with: (1) 27731 Gill Radiation Shield Hex Plug (1) Calibration Card (1) Resource D

Strona 36

Model HC2S3 Temperature and Relative Humidity Probe Temperature Accuracy over Temperature: 2.2 Relative Humidity Sensor Sensor: ROTRONIC Hygromer®