rubber seals on the ends of the shroud are clean and in good condition.
Position one end of the shroud over the lower snout and twist the top part
of the shroud, allowing it to extend and cover the upper snout; while doing
so, avoid making contact with the sonic transducers. Continue twisting the
shroud until it is fully extended and covering both snouts. Twist the
shroud so that the gas tubes and temperature thermistor cable are directed
towards the back of the sensor. Hang the tubes and cable over the trunk of
the sensor to alleviate any strain on the optical arms. See Figure 9-3, The
zero-and-span shroud mounted on the zero-and-span stand.
Intake
Thermistor Cable
Exhaust
FIGURE 9-3. The zero-and-span shroud mounted on the zero-and-
span stand. The top tube is the gas intake, the black cable is the
temperature thermistor, and the short bottom tube is the gas
exhaust.
4. Disconnect the IRGASON temperature sensor from the EC100 and
connect the shroud temperature sensor in its place.
5. Connect the EC100 to a PC with the EC100 USB cable (p/n 26561).
6. Resume power to the EC100. Wait for all the Gas and Power LED status
lights on the EC100 panel to turn green.
7. Launch ECMon, select the appropriate USB port, and click Connect. The
main screen should now be reporting real-time CO
2
and H
2
O
concentrations. Click Zero/Span. A graph will appear in the lower half
of the zero-and-span window showing measured CO
2
and H
2
O
concentrations (see Figure 9-4).