Met One 034B WindsetRevision: 1/11NCopyright © 1980-2011Campbell Scientific, Inc.

Strona 2 - Warranty and Assistance

Met One 034B Windset TABLE 5-1. Wind Speed Multiplier* Model Meters/Second Miles/Hour 034B M = 0.7989 Off = 0.28 M = 1.787 Off = 0.63 *When con

Strona 3

Met One 034B Windset 5.4 Example Programs The following programs measure the 034B every 5 seconds, and store mean wind speed, unit vector mean direct

Strona 4 - Figures

Met One 034B Windset 5.4.2 CR10X Example Program ;{CR10X} *Table 1 Program 01: 5.0000 Execution Interval (seconds) 1: Pulse (P3) 1: 1 Reps

Strona 5

Met One 034B Windset 10: Set Active Storage Area (P80) 1: 1 Final Storage Area 1 2: 101 Array ID 11: Real Time (P77) 1: 122

Strona 6 - 3. Installation

Met One 034B Windset 5.5.1 Sample CR10(X) Program when Long Leads are Required ;{CR10X} ; *Table 1 Program 01: 10 Execution Interval (seconds) 0

Strona 7

Met One 034B Windset 08: Wind Vector (P69) 1: 1 Reps 2: 0 Samples per Sub-Interval 3: 00 S, θu, & σ(θu) Polar*** 4: 1* Wind Spe

Strona 8 - 4. Wiring

Met One 034B Windset 12 Symptom: Incorrect wind direction 1. Verify that the Excitation voltage, Range code, multiplier and offset parameters are cor

Strona 9 - 5. Programming

Appendix A. Wind Direction Sensor Orientation A.1 Determining True North and Sensor Orientation Orientation of the wind direction sensor is done aft

Strona 10 - 5.2 Wind Direction

Appendix A. Wind Direction Sensor Orientation Subtract declination from 360° Add declination to 0° 22 E18 E16 E14 E12 E10 E8 E6 E4 E2 E02

Strona 11 - 5.4 Example Programs

Appendix A. Wind Direction Sensor Orientation FIGURE A-3. Declination Angles West of True North Are Added to 0 to Get True North A-3

Strona 12 - 5.4.2 CR10X Example Program

Warranty and Assistance The MET ONE 034B WINDSET is warranted by Campbell Scientific, Inc. to be free from defects in materials and workmanship under

Strona 13 - 5.5 Long Lead Lengths

Appendix A. Wind Direction Sensor Orientation A-4

Strona 14

Appendix B. Wind Direction Measurement Theory It is not necessary to understand the concepts in this section for the general operation of the 034B Wi

Strona 15 - 7. Troubleshooting

Appendix B. Wind Direction Measurement Theory B-2

Strona 16 - 8. References

Appendix C. Met One Instruments’ 034B Operation Manual This appendix contains a copy of Met One Instruments’ 034B Operation Manual. Campbell Scienti

Strona 17 - Orientation

Appendix C. Met One Instruments’ 034B Operation Manual C-2

Strona 18

Appendix C. Met One Instruments’ 034B Operation Manual C-3

Strona 19

Appendix C. Met One Instruments’ 034B Operation Manual C-4

Strona 20

Appendix C. Met One Instruments’ 034B Operation Manual C-5

Strona 21 - Measurement Theory

Appendix C. Met One Instruments’ 034B Operation Manual C-6

Strona 22

Appendix C. Met One Instruments’ 034B Operation Manual C-7

Strona 23 - 034B Operation Manual

Met One 034B Table of Contents PDF viewers note: These page numbers refer to the printed version of this document. Use the Adobe Acrobat® bookmarks

Strona 24

Appendix C. Met One Instruments’ 034B Operation Manual C-8

Strona 25

Appendix C. Met One Instruments’ 034B Operation Manual C-9

Strona 26

Appendix C. Met One Instruments’ 034B Operation Manual C-10

Strona 27

Appendix C. Met One Instruments’ 034B Operation Manual C-11

Strona 28

Appendix C. Met One Instruments’ 034B Operation Manual C-12

Strona 29

Appendix C. Met One Instruments’ 034B Operation Manual C-13

Strona 30

Appendix C. Met One Instruments’ 034B Operation Manual C-14

Strona 31

Appendix C. Met One Instruments’ 034B Operation Manual C-15

Strona 32

Appendix C. Met One Instruments’ 034B Operation Manual C-16

Strona 33

Appendix C. Met One Instruments’ 034B Operation Manual C-17

Strona 34

Met One 034B Table of Contents ii Figures 3-1. 034B Mounted on a Crossarm Using a 17953 NU-RAIL Crossover Fitting ...

Strona 35

Appendix C. Met One Instruments’ 034B Operation Manual C-18

Strona 37

Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbel

Strona 38

Met One 034B Windset 1. General The 034B Windset is used to measure horizontal wind speed and direction. Wind speed is measured with a three cup ane

Strona 39

Met One 034B Windset Wind Direction Measurement Range: 0 to 360° Threshold: 0.4 m s-1 (0.9 mph) Accuracy: ±4° Resolution: 0.5° Potentiometer Resis

Strona 40

Met One 034B Windset The set screw holes must be covered with the labels to prevent corrosion and assure the warranty. CAUTION Mount the CM200-series

Strona 41

Met One 034B Windset CM220 CM200 Series Crossarm FIGURE 3-2. CM200 Series Crossarm with CM220 Right Angle Mounting Bracket Attach the sensor cable t

Strona 42

Met One 034B Windset The CR10X, CR23X, and dataloggers programmed with CRBasic can also measure wind speed on a control port. With this option the bl