Model HFP01SC Self-Calibrating Soil Heat Flux Plate
;{CR10X}
;
*Table 1 Program
01: 1 Execution Interval (seconds)
;Measure HFP01SC on smallest range.
;
1: Volt (SE) (P1)
1: 6 Reps
2: 22 7.5 mV 60 Hz Rejection Range
3: 1 SE Channel
4: 7 Loc [ shf_mV_1 ]
5: 1 Mult
6: 0 Offset
;Measure HFP01SC on larger range.
;
2: Volt (SE) (P1)
1: 6 Reps
2: 23 25 mV 60 Hz Rejection Range
3: 1 SE Channel
4: 44 Loc [ shf_mV_1a ]
5: 1 Mult
6: 0 Offset
;Load in the factory calibration.
;
3: If (X<=>F) (P89)
1: 13 X Loc [ cal_1 ]
2: 1 =
3: 0 F
4: 30 Then Do
;Factory calibration in W/(m^2 mV) = 1000/sensitivity.
;
4: Z=F (P30)
1: 1 F ;<- Enter the unique calibration for plate 1 here.
2: 0 Exponent of 10
3: 13 Z Loc [ cal_1 ]
5: Z=F (P30)
1: 1 F ;<- Enter the unique calibration for plate 2 here.
2: 0 Exponent of 10
3: 14 Z Loc [ cal_2 ]
6: Z=F (P30)
1: 1 F ;<- Enter the unique calibration for plate 3 here.
2: 0 Exponent of 10
3: 15 Z Loc [ cal_3 ]
7: Z=F (P30)
1: 1 F ;<- Enter the unique calibration for plate 4 here.
2: 0 Exponent of 10
3: 16 Z Loc [ cal_4 ]
8: Z=F (P30)
1: 1 F ;<- Enter the unique calibration for plate 5 here.
2: 0 Exponent of 10
3: 17 Z Loc [ cal_5 ]
9: Z=F (P30)
1: 1 F ;<- Enter the unique calibration for plate 6 here.
2: 0 Exponent of 10
3: 18 Z Loc [ cal_6 ]
10: End (P95)
18