CDM-VW300 Series Dynamic Vibrating-Wire Analyzers
7.4.2 Data-Acquisition System Power
Power supply requirements will vary depending on system configuration and
location. The following power supplies are typically adequate to power a data-
acquisition system consisting of one datalogger, one CDM-VW300, and the
SC-CPI interface:
PS100 connected to ac
PS200 connected to ac
CR3000 rechargeable base connected to ac
CR3000 with alkaline base is NOT an adequate power supply
and will fail shortly after deployment.
NOTE
A solar powered system requires careful planning including a power budget.
For example, to power a data-acquisition system consisting of a CR3000
datalogger, an SC-CPI interface, and a CDM-VW300 series analyzer, the
following battery and solar panel specifications are required to operate the
system year-round in a remote installation in northern Utah using standard
silicon cell solar panels, lead-acid rechargeable batteries, and industry accepted
reserve margins:
CDM-VW300, two channels enabled
o 60 watts of solar panel capacity
o 91 ampere-hours of battery reserve
CDM-VW305, eight channels enabled
o 84 watts of solar panel capacity
o 130 ampere-hours of battery reserve.
Many applications will require the use of several analyzers. A carefully
considered power budget in these applications is essential to ensuring system
reliability and data continuity. Contact a Campbell Scientific application
engineer for assistance in configuring an adequate power supply.
7.5 Configuring the CDM-VW300 Analyzer
CDM-VW300 series analyzers must be configured by the user before sensor
measurement will be successful. Configuration can be done using DVWTool or
DevConfig software or a datalogger CRBasic program. See Section 7.1, PC
Based Tools, for more information about the software. See Section 7.10,
CRBasic Programming, for more information about using the datalogger.
IMPORTANT — Do not connect the CDM-VW300 analyzer or SC-CPI
interface to a PC until AFTER installing DVWTool 1.0 or later or DevConfig
2.04 or later. Consult Section 7.1.1, Software and Driver Installation, for more
information.
32