CRS451/CRS456 Submersible Water-Level Recording Sensor
The barometric pressure sensor data are shown below, also in blue.
8. Maintenance
Campbell Scientific recommends that the CRS451/CRS456 be factory
recalibrated every 24 months. Before a CRS451/CRS456 is sent to Campbell
Scientific, the customer must get an RMA (returned material authorization)
number and fill out the Declaration of Hazardous Material and
Decontamination form.
8.1 Every Visit
Collect data.
Visually inspect the physical conditions.
Check battery condition by using a laptop and HydroSci software to view
the battery voltage (see Section 6.6, Battery Voltage Reading). The
battery replacement procedure is provided in Section 8.3, Battery
Replacement.
Check all readings; adjust offsets if necessary.
Check recent data.
8.2 Every Two to Three Years or on a Rotating Schedule
Send the CRS451/CRS456 in for inspection and calibration.