Instruction Manual
5
4.2 5-pin Screw Terminal Plug Connector
The datalogger connects to the CS100 via a 5-pin screw terminal plug
connector. This connector is removable and may be replaced. The
replacement connector may come with a connector key attached to it to
ensure that the connector is plugged into the CS100 right side up (see
Figure 3). When the connector is right side up, it will easily plug into the
barometer.
Figure 3. Connector Key Attached to
5-pin Screw Terminal Plug Connector
A 5-pin screw terminal that is plugged in upside
down will damage the sensor—perhaps beyond
repair.
5. Programming
The CS100 sensor is measured using the singled-ended voltage measurement
instruction.
Atmospheric pressure changes little with time. In most weather station
applications measuring pressure once an hour is adequate.
Program Example 1 is for the dataloggers that use the CRBasic language,
such as CR200, CR800, CR850, CR1000, CR3000, CR5000, and
CR9000(X). In the example, the CR1000 measures the CS100 once an hour
in a program that runs at 1 Hz. In order to keep the CR1000 running in a
pipeline mode, the measurement instruction is placed outside the “If”
statement. The measurement is made every scan, and the measured value is
first written into a temporary variable called "CS100_temp". Once the
CS100 is turned on one minute before the hour, the CS100 starts to make the
correct pressure measurements. At the top of the hour, the correct value is
copied into the current variable called "pressure", and the sensor is turned off
immediately.
WARNING