DMM600 Duff Moisture Meter
12
11. Repeat steps 2 through 8.
12. Repeat steps 9 through 11 for other target volumetric water contents to
provide a range of calibration water contents.
13. Derive the calibration equation using polynomial curve fitting such as the
trendline function of Microsoft Excel.
7. Maintenance
7.1 Changing the 9-Volt Battery
The 9-volt battery is installed behind the display. To replace the battery,
remove the 3 outer screw from the display end cap and slide out of the main
cylinder. A screwdriver or small coin can be inserted into small slots on the
edge of the cylinder to assist separation.
8. Drying Samples and Calculating Volumetric Water
Content
8.1 Conventional Oven Method
For this method a thermostatically controlled oven is required and sample
containers suitable for oven temperature of 105°C.
Procedure
1. Weigh and record the weight of sample container
2. Place the sample in the container and record the weight of the container
and sample.
3. Place the container with sample in the drying oven and dry at 105°C for 12
to 16 hours or until sample weight does not decrease with drying time.
4. Remove sample from oven and allow the material and container to cool to
room temperature. Determine and record the weight of the dried sample
and the container using the same balance.
Calculation of gravimetric water content of the sample using spreadsheet or as
follows:
water content
massof container and wet sample massof container and dry sample
massof containerand drysample massof container
=
−
−
*100
Multiply gravimetric water content by bulk density to obtain volumetric water
content.