Model HFP01SC Self-Calibrating Soil Heat Flux Plate
If ( IfTime (4,CAL_INTERVAL,Min) ) Then
Move (shf_mV_180(1),4,shf_mV_run(1),4)
Move (V_Rf_180(1),4,V_Rf_run(1),4)
sw12_state = FALSE
EndIf
If ( IfTime (END_CAL,CAL_INTERVAL,Min) ) Then
Move (shf_mV_end(1),4,shf_mV_run(1),4)
'Compute new HFP01SC calibration factors.
For j = 1 to 4
shf_cal(j) = (V_Rf_180(j)*V_Rf_180(j)*128.7)/ABS (((shf_mV_0(j)+shf_mV_end(j))/2)-shf_mV_180)
Next j
shf_cal_on = FALSE
EndIf
EndSub
BeginProg
'HFP01SC factory calibration in W/(m^2 mV) = 1000/sensitivity.
shf_cal(1) = HFP01SC_CAL_1
shf_cal(2) = HFP01SC_CAL_2
shf_cal(3) = HFP01SC_CAL_3
shf_cal(4) = HFP01SC_CAL_4
Scan (100,mSec,3,0)
'Measure the HFP01SC soil heat flux plates.
VoltDiff (shf_mV(1),4,mV50C,11,TRUE,200,250,1,0)
'Apply calibration to HFP01SC soil heat flux plates.
For j = 1 to 4
shf(j) = shf_mV(j)*shf_cal(j)
Next j
'Measure voltage across the heater (Rf_V).
VoltSe (V_Rf(1),4,mV5000,33,TRUE,200,250,0.001,0)
'Maintain a 100 sample running average.
AvgRun (shf_mV_run(1),4,shf_mV(1),100)
AvgRun (V_Rf_run(1),4,V_Rf(1),100)
CallTable (mean)
Call hfp01sc_cal
NextScan
EndProg