Model HFP01SC Self-Calibrating Soil Heat Flux Plate
TABLE 3. Wiring for Example 1
Description Color CR10(X)
Sensor Signal White 1H
Sensor Signal Reference Green AG
Shield Clear G
Heater Resistor Signal Yellow 1L
Heater Resistor Signal Reference Purple AG
Shield Clear G
Power Red SW12
Power Reference Black G
External Power Control jumper wire SW12-CTRL to C8
Example 1. Sample CR10(X) Program Using a Single-Ended Measurement Instruction
;{CR10X}
;
*Table 1 Program
01: 1 Execution Interval (seconds)
;Measure HFP01SC on smaller range.
;
1: Volt (SE) (P1)
1: 1 Reps
2: 22 7.5 mV 60 Hz Rejection Range
3: 1 SE Channel
4: 2 Loc [ shf_mV ]
5: 1 Mult
6: 0 Offset
;Measure HFP01SC on larger range.
;
2: Volt (SE) (P1)
1: 1 Reps
2: 23 25 mV 60 Hz Rejection Range
3: 1 SE Channel
4: 8 Loc [ shf_mV_a ]
5: 1 Mult
6: 0 Offset
;Load in the factory calibration.
;
3: If (X<=>F) (P89)
1: 3 X Loc [ cal ]
2: 1 =
3: 0 F
4: 30 Then Do
5