Section 3. Operations
4. A file for logging data was created when you started the OBS-3A Utility.
You can review data at any time with
Open and import the log file
directly into an Excel spreadsheet for post-survey processing and plotting
(see Section 3.18—Excel Spreadsheets)
3.12 Cyclic Sampling
This mode is for logging data at regular time intervals such as 1, 10, 15, 30,
etc. minutes for example.
1. Request Barometric Correction from the OBS menu. Be sure to do this
while the OBS-3A is at the surface. Doing so when the instrument is
submerged will result in large depth errors.
2. Click
and select sensors, statistics, depth units (meters or feet), water
temperature, and salinity. Selection of temperature and salinity only
affects the depth calculation. It does not influence temperature or salinity
measurements.
3. Configure the Wave Setup if you want to measure wave heights and
periods (see Section 3.8—Definitions). Do this before scheduling the
other sample parameters.
4. Select Interval, Duration, Rate, and Power level; see recommendations
in “Sampling Schedules” (Section 3.9). The duration must be longer than
the Record Length. The minimum duration for the Record Length will
be computed and displayed by the OBS-3A Utility.
5. Click Start Sampling to begin logging data. Unplug test cable; install
dummy plug and locking sleeve. The instrument is ready for deployment.