CR10 and VW Sensors, 2002/12/18 2
Instruction 28
with AVW1 Interface
Instruction 28 1. Use instruction 28 to excite the sensor. It returns (kHz)
2
.
04 - Starting Freq (units = 100 Hz): Enter a frequency below
the lowest frequency in the test data on the calibration sheet
(e.g. 14 for 1400 Hz). See Sensors and Sweeps, page 5.
05 - End Freq (units = 100 Hz): Enter a frequency higher than
the highest frequency in the test data on the calibration sheet
(e.g. 35 for 3500 Hz). See Sensors and Sweeps, page 5.
06 - No. of Cycles: Enter 250 (typical). See Sensors and
Sweeps, page 5.
09 - Mult: Enter 1.0
10 - Offset: Enter 0
2. Use instruction 39 to obtain the square root of the reading
obtained with instruction 28. The value will now be kHz.
3. (Optional) Use instruction 55 to apply polynomial factors to
convert the readings to engineering units. The “manual ABC”
listed on the sensor calibration record are intended to operate
on Hz readings, so they must be adjusted for the kHz value
returned by instruction 39 above. See below.
04 - C0: Enter the C factor listed in the calibration record for
the C0 parameter. Do not modify this value.
05 - C1: Multiply the B factor by 1,000 (move the decimal
three places to the right) and enter the resulting value for the
C1 parameter.
06 - C2: Multiply the A factor by 1,000,000 (move the decimal
six places to the right) and enter the resulting value for the C2
parameter.
07 - C3: Enter 0.
08 - C4: Enter 0.
08 - C5: Enter 0.