Temperature Zero Readings & Thermal Factors
NOTE: TEMPERATURE ZERO AND THERMAL FACTORS ARE USED FOR TEMPERATURE
CORRECTION. TEMPERATURE CORRECTION MAY BE USED TO COMPENSATE FOR CHANGES IN
A GAGES READINGS DUE TO TEMPERATURE CHANGE. TEMPERATURE EFFECT ON READINGS
VARIES ON DIFFERENT MODELS OF GAGES AND TEMPERATURE CORRECTION IS OFTEN NOT
REQUIRED. DEPENDING ON THE GAGE MODEL TEMPERATURE CORRECTION MAY NOT BE
AVAILABLE AND SOME GAGE MODELS REQUIRE A DIFFERENT FORMULA TO COMPENSATE.
TEMPERATURE CORRECTION IS COMMENTED OUT IN THE MAIN PROGRAM FILE BY DEFAULT.
Default values for Temperature Zero Readings and Thermal Factors are 0. Temperature
Zero Readings and Thermal Factors can be added if the user requires the Datalogger to
use temperature compensation.
Navigate to the Main(.CR8/.CR1) file C:\Campbellsci\CRBasicEditor. Open the
Main(.CR8/.CR1) file with notepad and uncomment the line specified in the code for all
model gages that temperature correction is required.