26
APPENDIX F –8032 MULTIPLEXER Manual
The following Appendix includes excerpts from the Model 8032 Multiplexer manual that apply
to the Micro-800. A complete version of the Model 8032 Multiplexer Manual can be found at
WWW.GEOKON.COM.
THEORY OF OPERATION
The Model 8032 Multiplexer expands the number of channels that can be read by the
MICRO-800 Datalogger or MICRO-1000 Datalogger. Channel switching is accomplished
by mechanical relays mounted on the underside of the circuit board and the transducer
connections are accomplished by friction locking spring-loaded terminals mounted on the
top side of the circuit board.
Two switching configurations are supported, 16 channels of 4 conductors or 32 channels
of 2 conductors. For the 8032-C, these configurations are set by a DipSwitch on the top
side of the circuit board. A second DipSwitch selects whether the 8032-C is being used
with a datalogger or GK-403 Readout Box
To protect against lightning or EMI/RFI induced transients, each channel is protected by
an integrated lightning protection system, incorporating 230V tripolar plasma surge
arrestors, 150V bipolar plasma surge arrestors, 10uH inductors and 16V transient voltage
protection diodes. See Appendix A for complete specifications on these components.