Appendix E. Status Table
The AVW200 status table contains system operating status information
accessible via DevConfig, terminal emulator, or another PakBus device such as
a datalogger. Status Table information is easily viewed by going to DevConfig |
AVW200| Connect | Data Monitor | Status. The status table can be viewed via a
terminal emulator and command 4.
The status information can be retrieved by the datalogger by using the CRBasic
GetVariable instruction. Following is an example of retrieving the BattVoltage
status of the AVW200 using the CRBasic GetVariables instruction:
Public RC,AVW_BV
GetVariables(RC,ComSDC7,200,200,0000,0,"Status","BattVoltage",AVW_BV,1)
DevConfig polls the status table at regular intervals, updating
status information.
TABLE E-1 is a comprehensive list of status table variables with brief
descriptions.
TABLE E-1. Status Fields and Descriptions
Status
Fieldname
Description
User can change?
Record No Record number for this set of data No
Time Stamp Time the record was generated No
OSversion Version of the Operating System No
OSdate Date OS was released No
ProgName Name of the running program No
ProgSig Signature of the running program No
StationName User defined Station Name Yes
Compileresults Compile results of the running program Yes
PakBusAddress AV200 PakBus address Yes
RfInstalled Specifies the model number of the MaxStream radio
if it is recognized by the datalogger. It will have a
value of zero if there is no radio recognized by the
AVW200.
No
RfNetAddr Specifies the radio network address of the built in
radio. This setting should be set to match the
network address for the RF401 base used to
communicate with the datalogger.
Yes
RfNetHopSeq Specifies the hopping sequence that will be used for
the built-in radio. This value should be set to match
the value of the same setting for the RF401 base
station used to communicate with this datalogger.
Yes