
TDR100
19
6.2 Discussion of TDR Instruction Parameters (Instruction 119)
6.2.1 Parameter 1: SDM Address
The SDM address of the TDR100 is set by selecting hexadecimal values with
the thumbwheel switch on the TDR100 front panel. Instruction 119 requires a
2 digit integer that is the base 4 value of the TDR100 address. See Table 5-1
6.2.2 Parameter 2: Output Option
6.2.2.1 Enter 0: Measure La/L
In the section on TDR Principles, the equation
K
a
a
=
is presented. The terms of this equality can be related to volumetric water
content using empirical relationships such as equations [3] and [4] shown in
Section 7. The apparent length of the probe rods, L
a
, changes with water
content. Dividing L
a
by the real rod length, L, gives the square root of
dielectric constant. The L
a
/L value is empirically related to volumetric water
content using calibration functions of the form θ
v
= f(L
a
/L). See Section 7 for
commonly used calibration functions.
6.2.2.2 Enter 1: Collect Waveform
The digitized reflection waveform can be collected from the TDR100 by the
datalogger in ASCII format. Datalogger input storage must be manually
allocated for the waveform data points using “block allocation” in Edlog
datalogger software. The waveform is collected as an array. The first 9 cells
of the array is a header and contain values entered in Instruction 119.
Waveform values fill the rest of the array. Always allocate 9 more cells in
datalogger input storage than the number of datapoints chosen in parameter 6.
See Section 6.2.11.1 for header information.
6.2.2.3 Enter 2: Collect Waveform and First Derivative
The reflection waveform and the first derivative are written to datalogger input
storage. Input storage must be manually allocated. See Section 6.2.6. The
required number of input locations is twice the number of points (instruction
Instruction 119, parameter 6) plus 9 for the header information.
6.2.2.4 Enter 3: Measure Bulk Electrical Conductivity
The electrical conductivity routine in the datalogger returns the value
11
1Z
c
+
ρ
which must be multiplied by the probe constant, K
p
, to get an electrical
conductivity value. Kp is entered in Instruction 119 parameter 12. See
Komentarze do niniejszej Instrukcji