November 2005, ARM-TR-069 Appendix B Six-Month Procedures

Strona 3

November 2005, ARM-TR-069 B-1SURTHREF Six-Month Calibration Procedures Frequency: Twice a year Mentor: Michael Ritsche, Barry Lesht To verify

Strona 4

November 2005, ARM-TR-069 B-2

Strona 5

November 2005, ARM-TR-069 B-3SURFACE TEMPERATURE AND HUMIDITY REFERENCE SYSTEM SENSOR VERIFICATION Tested by:______________________ Date: _____

Strona 6

November 2005, ARM-TR-069 B-42. ROTRONIC MPH100H Temperature and RH (The temperature readings, once stabilized should match within +/- 0.5°C)

Strona 7

November 2005, ARM-TR-069 Appendix C Sensor Replacement Procedures

Strona 8 - Bi-Weekly Procedures

November 2005, ARM-TR-069 C-1Sensor Replacement Procedures Both the Rotronic and Vaisala probes have removable sensing units. To access the sensi

Strona 9

November 2005, ARM-TR-069 C-2 Once the service has been stopped, go to “Start” and choose “LoggerNet” to start the program. When the LoggerNet me

Strona 10 - Six-Month Procedures

November 2005, ARM-TR-069 C-3****************DAILY OUTPUT and LOGISTICS*************************** ; ;This output, once per day, captures the seria

Strona 11 - SURTHREF

November 2005, ARM-TR-069 C-4;V1 probe serial number 25: Z=F x 10^n (P30) 1: 0012 F 2: 0 n, Exponent of 10 3: 31 Z Loc [ V1probeSN ]

Strona 12 - November 2005, ARM-TR-069

November 2005, ARM-TR-069 Contents 1. Purpose...

Strona 13 - SENSOR VERIFICATION

November 2005, ARM-TR-069 C-5;R1 probe location 32: Z=F x 10^n (P30) 1: 4 F 2: 0 n, Exponent of 10 3: 38 Z Loc [ R1locate ] ;R2 p

Strona 14

November 2005, ARM-TR-069 C-6program that was just updated above (C:\SurTHRefPgm) and send it to the logger, click on “Ok” when the warning banner

Strona 15 - Sensor Replacement Procedures

November 2005, ARM-TR-069 Appendix D LoggerNet Set-Up

Strona 16

November 2005, ARM-TR-069 D-1LoggerNet Setup Setup and start PC. Ensure the PC is connected to the network and the loggers and NL-100’s are conn

Strona 17

November 2005, ARM-TR-069 D-2 Once the IPPort has been added, enter the IP address 198.124.98.19. Ensure the Communications Enabled box is checke

Strona 18

November 2005, ARM-TR-069 D-3 Once the IPPort has been added you can change the name. Suggestion would be to change the name to IPPort.Sur.19. T

Strona 19

November 2005, ARM-TR-069 D-4 Once the logger has been added you should rename it to SurTHRef either using the Rename button or highlighting and u

Strona 20

November 2005, ARM-TR-069 D-5Ensure the Communications Enabled box is checked and leave all settings under hardware tab as default. Click on the s

Strona 21

November 2005, ARM-TR-069 D-6 Click on the Clock tab to make changes to the following settings: Leave Time Zone Offset as 0h 00m. Under Automat

Strona 22 - Appendix D

November 2005, ARM-TR-069 D-7 You will not need to make any changes to the settings in the Final Storage Area 2 tab. Just ensure that the Enabled

Strona 23 - LoggerNet Setup

November 2005, ARM-TR-069 1Surface Temperature Humidity Reference System 1. Purpose The Surface Temperature and Humidity Reference (SURTHREF) S

Strona 24

November 2005, ARM-TR-069 D-8 The logger you added in the above steps will be located in the Stations menu. Highlight the logger you wish to conn

Strona 25

November 2005, ARM-TR-069 Appendix E Perl Script Set-Up

Strona 26

November 2005, ARM-TR-069 E-1PERL SCRIPT SET-UP ${DataDir} = The name of the output directory used by LoggerNet ${DataFile} = The name of the out

Strona 27

November 2005, ARM-TR-069 E-2 8. Click on Add Scheduled button on lower left hand side of window. 9. Rename the task in the window by right cli

Strona 28

November 2005, ARM-TR-069 E-3 Working Directory: C:\SurTHRefData The generic format is: C:\${DataDir} When finished, click on the Appl

Strona 29

November 2005, ARM-TR-069 Appendix F NL-100 Set-Up

Strona 30

November 2005, ARM-TR-069 F-1NL-100 SET-UP In order to communicate with the NL-100 you need a Null-modem cable and a PC or Laptop with a terminal

Strona 31 - Perl Script Set-Up

November 2005, ARM-TR-069 F-2 2. Strike the Enter button a few times to get the attention of t

Strona 32 - PERL SCRIPT SET-UP

November 2005, ARM-TR-069 Appendix G Wiring Diagrams

Strona 33

November 2005, ARM-TR-069 G-1Rotronics MP100H-1 wiring table (three sensors) Campbell CR23X Connection Wire Color Function Power Out – 12V G

Strona 34

November 2005, ARM-TR-069 2• Campbell Scientific NL-100 Ethernet Device. • Allied Telesyn AT-MC13 Media Converter. • Dell Computer. The com

Strona 35 - NL-100 Set-Up

November 2005, ARM-TR-069 G-2 Rotronics MP100H-1 sensor units Sensor Identifier Sensor Serial Number R1 92781 R2 92782 R3 92783 R4 92784 (spa

Strona 36 - baud: 115200

November 2005, ARM-TR-069 G-3

Strona 37

November 2005, ARM-TR-069 G-4 AC Power Wiring Diagram

Strona 38 - Wiring Diagrams

G-5 November 2005, ARM-TR-069 1 2 3 4 5 6 H L G H L H L G H L G H L G H L G E1 E2 E3

Strona 39

G-6 November 2005, ARM-TR-069

Strona 40

November 2005, ARM-TR-069 32. Inform the instrument mentor if the troubleshooting techniques in this manual or from experience are unable to loca

Strona 41

November 2005, ARM-TR-069 4Vaisala 1-888-824-7252 (follow prompts) Rotronic 1-800-628-7101 ext. 20 8. Shipment Notification All returns will be

Strona 42

November 2005, ARM-TR-069 5Every Six Months The following are required to perform the six-month checks, maintenance, and sensor verification (see

Strona 43

November 2005, ARM-TR-069 Appendix A Bi-Weekly Procedures

Strona 44

November 2005, ARM-TR-069 A-1SURTHREF Preventative Maintenance Procedures Frequency: Bi-Weekly Mentor: Michael Ritsche, Barry Lesht The fo