November 2005, ARM-TR-069
B-4
2. ROTRONIC MPH100H Temperature and RH
(The temperature readings, once stabilized should match within +/- 0.5°C)
(The RH readings, once stabilized should match within +/- 3% RH for 0 – 90% RH and
+/- 4% for 90 – 100% RH)
R1 MPH100H Serial Number________________
Temperature:
HMI31 Reading:___________R1 HMP-45D Reading:_____________ P/F
RH%:
HMI31 Reading:___________R1 HMP-45D Reading:_____________ P/F
R2 MPH100H Serial Number________________
Temperature:
HMI31 Reading:___________R2 HMP-45D Reading:_____________ P/F
RH%:
HMI31 Reading:___________R2 HMP-45D Reading:_____________ P/F
R3 MPH100H Serial Number________________
Temperature:
HMI31 Reading:___________R3 HMP-45D Reading:_____________ P/F
RH%:
HMI31 Reading:___________R3 HMP-45D Reading:_____________ P/F