6. Record the serial number. Write down the serial number of the new sensor and the date
and time it was installed on the Sensor Installation/Calibration Form.
7. Send the sensor head out for calibration. Complete a Sensor Calibration Metadata
Form (Appendix A.5) for each sensor that is sent out for calibration. Mail the sensor head
back to the manufacturer for recalibration. When the sensor comes back, fill in the rest of
the Sensor Calibration Metadata Form and keep the sensor as a duplicate.
8. Submit calibration information to the TEAM portal. Submit forms to the TEAM
portal as soon as technically feasible.
4.18 Uploading calibration forms
Forms needed:
Completed Sensor Installation/Calibration Form
Completed Sensor Calibration Metadata Form
After sensors have been removed from the field and sent out for calibration, the Sensor
Installation/Calibration Form and Sensor Calibration Metadata Form should have been
completed. To upload this information, follow the steps below:
a) Open an internet browser and navigate to the TEAM portal (www.teamnetwork.org).
b) Hold the mouse cursor over the “Products” link at the top of the page until a menu drops
down
c) From this menu, hold mouse over the “Data” link and select “Data Management” from
the next menu that appears.
d) Using the arrows on the right side of the screen, navigate to the “Climate Metadata” tab.
Click on this tab.
e) Choose your site from the dropdown menu and click “Go.”
f) Select the “Sensor Management” tab from the bottom of the page (see Figure 7 below)
and double-click in the “Status” column next to the sensor that is being calibrated (it
should say “Installed”).
g) If removing the sensor for calibration, choose the “Removed for Calibration” option from
the drop down menu. If you are sending the sensor out for calibration, select “Sent Out
for Calibration.” (Note: The status for a sensor can only be changed to “Sent Out for
Calibration” after its status has been changed from “Installed” to “Removed for
Calibration.”)
h) After changing the status of a sensor, a window will appear that should match the form
filled out in the field. Copy the field form information into this window and click
“Submit.”
Note: More detailed instructions for uploading calibration information can be found in Step 7 of
the Data Management Tool Help (Appendix A.7).