Section 7. Measurement Instructions
TCSE (Dest, Reps, Range, SEChan, TCType, TRef, MeasOfs, SettlingTime,
Integ, Mult, Offset)
S.E. Channel
Ground
Thermocouple
This instruction measures a thermocouple with a single-ended voltage
measurement and calculates the thermocouple temperature (°C) for the
thermocouple type selected. The instruction adds the measured voltage to the
voltage calculated for the reference temperature relative to 0° C, and converts
the combined voltage to temperature in °C.
Voltage and Thermocouple Parameters
Parameter
& Data Type
Enter
Dest Variable
or Array
The Variable in which to store the results of the instruction. When Reps are used the
results are stored in an array with the variable name. An array must be dimensioned to
have elements for all the Reps.
Reps
Constant
The number of repetitions for the measurement or instruction.
Range
Constant
Alpha Code
Voltage Range
mV5000
± 5000 mV
mV2500
± 2500 mV
mV250
± 250 mV
mV25
± 25 mV
mV7_5
± 7.5 mV
mV2_5
± 2.5 mV
Autorange
mV2_5 mV5000
Selects range (Sect. 3.1)
mV250C
± 250 mV The mV250C, mV25C, mV7_5C, and
mV25C
± 25 mV mV2_5C ranges pull the channel into
mV7_5C
± 7.5 mV common mode range and check for open
mV2_5C
± 2.5 mV input
AutorangeC
mV2_5 mV250
Selects C range
DiffChan
Constant
The differential channel number on which to make the first measurement. When Reps are
used, subsequent measurements will be automatically made on the following channels. If
the channel is entered as a negative number, all reps occur on that channel.
SEChan
Constant
The single-ended channel number on which to make the first measurement. When Reps
are used, subsequent measurements will be automatically made on the following channels.
If the channel is entered as a negative number, all reps occur on that channel.
TCType Alpha Code
Thermocouple Type
Constant TypeT
Copper Constantan
TypeE
Chromel Constantan
TypeK
Chromel Alumel
TypeJ
Iron Constantan
TypeB
Platinum Rhodium
TypeR
Platinum Rhodium
TypeS
Platinum Rhodium
TRef
Variable
The name of the variable that is the reference temperature for the thermocouple
measurements.