SECTION 9.
INPUT/OUTPUT
INSTRUCTIONS
TABLE
9-3. Thermocouple
Type Codes
Code
X1
x2
X3
x4
Thermocouple
Tvpe
T
(copper
-
constantan)
E
(chromel
-
constantan)
K
(chromel-
alumel)
J
(iron -
constantan)
Normal
Measurement
TC
input from A5B40
isolation
(use
5
V range)
X
=
9 Output
-99999
if out of
common mode
(lnst.
14 only)
X=0
X=8
TABLE
9-4i
Voltage and
Temperature
Ranges
for
Thermocouples if
Reference Junction is 20oC
Voltdge
Range
Temperature Range
oC
Type
T
Type E
Type K
Type
J
t5
t15
t50
t500
ilnV
inV
dnV
-122
-270
-270
-270
141
386
1254
1372
-131
to
-270
to
-270
to
113
294
760
to
to
to
to
-70
to 98
-270
to
237
-270
to 676
-270
to
1000
131
316
400
-84
to
-210
to
-210
to
combined
differential
V
***
14
THERMOCOUPLE
TEMPERATURE,
***
DIFFERENTIAL
MEASUREMENT
FUNCTION
This instructbn
calculates the
thermocouple
temperature
[or
the thermocouple
type
selected.
ibn specifies
a DIFFERENTIAL
VOLTAGE
REMENT
(Section
13.2)
on
the
adds the
measured
voltage to
the voltage
lated for the
reference
temperaturelrelative
to
OoC,
and converts the
Input locations altered:
1 for each
thermocouple
channel
***
16 TEMPERATURE
FROM
***
PLATINUM
R.T.D.
FUNCTION
This
instruction
uses the
result of a
previous
RTD
bridge
measurement to
calculate the
temperature according to
the
DIN 43760
specification adjusted
(1980)
to
conform
to
the
pending
International
Electrotechnical
Commission standard.
The
polynomial
range
is
-200oC
to
850oC.
The
error
is less than
0.001oC
between
-100
and
+300oC, and is
less than 0.003oC
between
-180oC
and +830oC.
The error
(T
calculated
-
T
standard)
is +0.006oC at
-200oC
and
-0.006oC
at
+850oC. The input must be the
ratio
Rr/Rs,
where R. is the
RTD resistance and
Rs the
resistance of the
RTD at OoC
(Sections
7.9
and
7.10).
PARAM. DATA
NUMBER
TYPE DESCRIPTION
Repetitions
Input
location of Rs/Ro
Input
location of result
Multiplier
Otfset
to
temperature
in
oC.
The
are
briefly
shorted
to
ground
prior
to
the
voltage measurement to
insure
that
tltev
are
within
the common
mode
range. Tablf
9-3
gives
the
thermocouple
type
codes for
4. Select
the smallest
voltage
rangle that
will
accommodate
the
temperaturet
being
measured
(Table
9-4).
PARAM.
DATA
NUMBER
TYPE DESCRIPTION
Repetitions
Range
code
First
TC channel
Thermocouple
type code
Reference
temperature
location
Destination
input location
Multiplier
Otfset
01:
02:
03;
o4:
05:
06:
07:
08:
2
2
2
2
4
2
4
4
FP
FP
01:
02:
03:
04:
05:
4
FP
FP
Input locations
altered:
1 for each RTD
9-7