SECTION 7.
MEASUREMENT
PROGRAMMING
EXAMPLES
The
manual for the
101 Probe
gives
the
coefficients of
the sth order
polynomial
used to
convert
the output in
millivolts to temperature
(E
denotes the
power
of 10 by
which the mantissa
is multiplied):
c0
-53.7842
c1
0.147974
c2
-2.18755E-4
c3
2.19046E-7
c4
-1
.1 1341E-10
c5 2.33651E-14
The 21X willonly
allow
5 significant
digits to
the
right
or left o{ the decimal
point
to be
entered
from the keyboard.
The
polynomialcannot
be
applied
exactly as
given
in
the 101
manual. The
initial millivolt reading
must be scaled
if the
coefficients of the higher order terms
are
to
be
entered with the maximum
number
of
significant
digits. lf 0.001
is used as a multiplier on
the
millivolt
output, the coefficients are
divided by
0.001 raised to
the appropriate
power (i.e.,
C0=C0,
C1
=C110.001,
C2=C2/.000001,
etc.).
With this adjustment,
the coefficients
entered
in
Parameters 4-9 of Instruction
55
become:
01:
P4
01: 5
02: 5
03:
1
04:.
1
05: 10
06:
2000
07:
1
08: .001
09: 0
PROGRAM
Excite, Delay,Volt(SE)
Reps
5000 mV
slow Range
lN Chan
Excite all
reps
w/EXchan
1
Delay
(units.01sec)
mV Excitation
Loc
:
Mult
Offset
c0
-53.784
c1 147.97
c2
-218.76
c3 219.05
c4
-111.34
c5
23.365
FIGURE 7.15-1.
101 Thermistor
Probes
Connected to 21X
02: P55 Polynomial
01: 5 Reps
02:
1
X Loc
.03:
1 F(X) Loc:
04:
-53.784
C0
05:
147.97
C1
06:
-218.76
C2
07: 219.05 C3
08:
-111.34
C4
09: 23.365 C5
7.16
MEASURING
107 TEMPERATURE
PROBE
WITH
60 HZ
REJECTION
The 107
temperature
probe
is normally
measured with Instruction 11 which uses the
same measurement sequence
as
lnstruction
5
with
a
fast
integration
voltage
measurement.
The fast
integration does not reject 60 Hz
noise.
In environments where 60
Hz
is a
problem,
the
107
probe
can be measured with
Instruction
5
using the
slow
integration time.
A multiplier of 800 in Instruction
5
is used to
scale the voltage measurement for the
polynomial
Instruction
55
which
is
used
to
convert the
voltage measurement
to
oC.
THERMISTOR
PROBES
EX1
+
Hr I
21X
LO1
Ht2
RED
LEADS
TO
SINGLE ENDED
INPUT
CHANNELS I-5
7-12